文章詳目資料

中原學報

  • 加入收藏
  • 下載文章
篇名 可攜式胎兒心電圖監視器之研究
卷期 26:3、26:3
並列篇名 The Development and Implementation of A Portable Fetal Electrocardiogram Monitoring System
作者 徐良育
頁次 63-69
關鍵字 胎兒心電圖訊號處理胸腹轉移函式數位訊號處理器Fetal ECGThoracic-abdominal transfer functionSigal processing
出刊日期 199808

中文摘要

     可攜式胎兒監視系統可幫助高危險群孕婦即早監視胎兒之不正常發展。一般而言,使用非侵入式的胎兒心電圖監視器是最合適且最方便的技術。可是,由母體腹部電極所取得的訊號仍受母親之心電圖、肌電圖及電源所干擾,而造成訊號處理上的困擾。本研究提出利用

英文摘要

     A portable non-invasive fetal electrocardiogram (FECG) monitoring systemis very helpful for high-risk pregnant women to monitor the grow of fetus.However, signals obtained from maternal abdomen are contaminated by maternal ECG(MECG), electromyogram (EMG), and interference from electric device. Correlationanalysis is commonly used to extract the FECG. However, the morphology andamplitude of ECGs are moderated by respiration and motion. Thus, largelow-frequency noises remain in the processed signal. A new signal processingmethod, specifically the thorax-abdomen transfer function, was developed in thislaboratory to overcome these drawbacks. By incorporating this method with adigital signal processing chip, a portable real-time FECG monitoring system isdeveloped. The system can not only process the FECG in real time but also storethe fetal heart rates for further analysis. Ten pregnant women were recruitedfrom VGH Taipei to evaluate this system. The FECGs that were obtained by thissystem are comparable with those processed by off-line method. On the otherhand, three sets of thoracic and abdominal ECGs in combination with foursimulated FECGs, the amplitudes are 1/4, 1/5, 1/7, and 1/10 of maternal QRS, areused to evaluate the accuracy and efficiency of this system. The resultindicated that the detection rate is in the range of 79.79 % to 99.68 %. Inaddition, it clearly demonstrated that the detection rate is highly correlatedwith the amplitude of FECG. This system is proven to be convenient and accuratefor long-term fetal monitoring.

相關文獻