篇名 Taguchi - Method Assessment of Brain Perfusion Nuclear Medicine Scan of the Best Tuning Parameter
卷期 24:3
並列篇名 應用田口方法評量核醫腦部灌注斷層掃描檢查之最佳調整參數
作者 Hsueh- Ying ChenCheng- Tao SieCheng- Tao SaiHsin- Yi WangWan-Yu Lin
頁次 123-131
關鍵字 Padionuclide Brain Perfusion TomographyTaguchi-methodSignal/nise ratio核醫腦部灌注斷層掃描田口方法信號/雜訊比
出刊日期 201109

中文摘要

目的:使用高解析平行孔洞式準直儀、應用田口方法評量核醫腦部灌注斷層掃描檢查之最佳調整參數。方法:本實驗在評估核醫腦部灌注斷層掃描檢查過程中,歸納出放射性製劑之活性強度、病患每一個角度平面影像之攝取時間與攝影掃描轉動角度等三種影響最大之控制因子;每一個控制因子皆使用三個變動水準,故採用田口方法中之L9(34)直交表並使用Hoffman腦部假體掃描來評估每組不同調整參數下的影像晶質。核醫腦部灌注斷 層掃描共9組27張影像優劣之評估,由二位核醫科專科醫師審視其影像對稱性及均句度給予分數,最後再經由信號/雜訊比與變異數分析統計分析結果。結果:依據田口方法望大特性S/Nratio '得到核醫腦部灌注斷層掃描之最佳化造影條件為:放射性製劑之活性強度2.5mCi、病患、每一個角度平面影像之攝取時間的秒與攝影掃描轉動角度3度。結論:應用田口方法評量核醫腦部灌注斷層掃描之最佳調整參數研究中,主要取決於造影時間及角度之選擇;由控制因子分類觀念得知,放射性製劑之活性強度對影像品質有一定影響力,但由於本實驗驗證中發現如使用1.8mCi仍可獲得相當品質之影像,故可經由調整此類控制因子來使放射性製劑成本降至最低。

英文摘要

Methods: Inthis study, the brain perfusion scintigraphy in the assessment of SPECT scan process, the sum intensity of the activity of radioactive agents,位le patient each time point of the intake plane image rotation angle and photo scanning con甘olof也e three most influential factors; each con仕'01 factorτbree levels ofvariations are used, So the use ofTaguchi - method ofL9 (34) 。rthogonalarray and use the Hoffman brain phantom to evaluate each of the different scan p缸缸neters to adjust the image quality. Brain SPECT scans nuclear medicine group of 27 files 9 assessment of 甜engthsand weaknesses, by two nuclear medicine physicians Examined the images to give scores of symme旬and uniform旬" and finally t尬。ugh the signal / noise ratio (SIN ratio) andAnalysis ofvariance (ANOVA) statistical analysis. Results: Taguchi method based on looking great features S / N ratio, by radionuclide brain perfusion imaging scans of the optimal conditions: the activity of radioactive pr叩紅ations 甜ength2.5 mCi, patient each time po血t of the intake of planar images of 45 seconds sc缸ming and photography Rotation angle of 3 degrees. Conclusion: Taguchi method assessment of brain perfusion nuclear medicine scan study of the optimal adjus恤ent of parameters, depending on the choice of imaging time and angle; classification conc叩ts learned by the control factor, intensity of activity of radioactive agents have an influence on image quality However, as found扭曲e course of the experinlent using 1.8mCi, such as the image quality will still be considerable, it can be adjusted to control factors such 也atthe radioactive prep前的ion costs to a minimum.

相關文獻